NFPA 99-2012: 10.2.3.6 Power strips providing power to patient care-related electrical equipment must be Special
purpose Relocatable Power Taps (SPRPT) listed as UL 1363A or UL 60601-1. Power strips providing power to non- patient-care-related electrical equipment must be Relocatable Power Taps (RPT) listed as UL 1363. Power strips may not be used in a patient care vicinity to power non-patient care-related electrical equipment (e.g., personal electronics) They must also meet the requirements of NFPA 99: 10.2.3.6. #ASHEComplianceTip